8. Detailed Data Setting

8. Detailed Data Setting

8.1 Current Output Setting

This section describes setting of the analog output range. Table 8.1 shows the parameter
codes for each setting item and set values.

Table 8.1 Display Items

T8.1.1.EPS

Set item

Parameter code Set value

Current output

C01

0: Oxygen concentration

1: Humidity

2: Mixing ratio

Output mode

C03

0: Linear

1: Logarithm

Min. oxygen concentration

C11

Oxygen concentration reading corresponding to 4 mA

Max. oxygen concentration

C12

Oxygen concentration reading corresponding to 20 mA

Min. humidity

C13

Humidity reading corresponding to 4 mA

Max. humidity

C14

Humidity reading corresponding to 20 mA

Min. mixing ratio

C15

Mixing ratio at 4 mA

Max. mixing ratio

C16

Mixing ratio at 20 mA

Output damping coefficient

C30

0 to 255 seconds

CAUTION

When you select logarithmic mode in Section 8.1.3, “Output Mode,” later in this
manual, the oxygen concentration, humidity reading, and mixing ratio remain constant at
0.1% O

2

, 0.1% H

2

O and 0.01 kg/kg respectively.