IM 11M12A01-05E

7-14

7.7.3

Minimum Current (4 mA) and Maximum Current (20 mA) Settings

This section describes how to set the humidity readings corresponding to 4 mA and 20
mA to 30% H

2

O and 80% H

2

O respectively.

Switch operation Display Description

ENT

A

01

Display after the password has been entered.

ENT

C

01

Set the humidity reading at 4 mA. Change the parameter

code to C13. Touch the [

] key to switch to Group C.

ENT

C

0

1

Touch the [

] key to move the position of the digit that is

flashing to the right.

ENT

C

1

1

Touch the [

] key to enter the number 1.

ENT

C1

1

Touch the [

] key to move the position of the digit that is

flashing to the right.

ENT

C1

3

Touch the [

] key to enter the number 3.

ENT

0

00

Touch the [ENT] key to display the current set value. The

humidity 0% H

2

0 is now being displayed.

ENT

0

0

0

Touch the [

] key to move the position of the digit that is

flashing to the right.

ENT

0

3

0

Touch the [

] key to enter the number 3.

ENT

030

If you touch the [ENT] key, all the digits flash.

ENT

030

Touch the [ENT] key again to stop flashing.

ENT

C

13

Touch the [ENT] key to return to the parameter code

selection display.

ENT

C1

3

Set the humidity reading at 20 mA. Touch the [

] key to move

the position of the digit that is flashing to the right.

ENT

C1

4

Touch the [

] key to change the number 3 in C13 to "4."

ENT

0

25

Touch the [ENT] key to display the current set value.

ENT

0

2

5

Touch the [

] key to move the position of the digit that is

flashing to the right.

ENT

0

8

5

Touch the [

] key to change the number 2 in C25 to "8."

ENT

08

5

Touch the [

] key to move the position of the digit that is

flashing to the right.

ENT

08

0

Touch the [

] key to change the number 5 in C85 to "0." The

number changes from 5 to 6 . to 9 to 0.

ENT

080

If you touch the [ENT] key, all the digits flash.

ENT

080

Touch the [ENT] key again to stop flashing.

ENT

C

14

If you touch the [ENT] key again, the parameter code

selection display will appear.

ENT Basic

Touch the [

] key together with the [ENT] key to return to the

panel basic panel display. (This is not required if you proceed to

display make another setting.)

(The displayed numeric characters indicate the humidity of

the measurement gas.)

T7701.EPS

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.