1 terminals for the external wiring, 2 wiring, Wiring procedure -4 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 65: Wiring power and ground terminals -5, Figure 5.1 terminals for external wiring, Figure 5.2 wiring connection

Advertising
background image

IM 11M12A01-05E

5-2

5.1.1

Terminals for the External Wiring

Remove the terminal cover on the opposite side of the display to gain access to the
external wiring terminals.

1

2

C

DO

DO

DI

1

2

L

N

FG

+

-

AO

G FG

F5.1E.EPS

Figure 5.1 Terminals for External Wiring

5.1.2

Wiring

Make the following wiring for the equipment. It requires a maximum of four wiring
connections as shown below.

(1) Analog output signal
(2) Power and ground
(3) Contact output
(4) Contact input

1

DI-1

2

DI-2

3

DI-C

4

DO-1

5

DO-1

6

DO-2

7

DO-2

8

FG

9

AO

(+)

10

AO

(-)

11

L

12

N

13

G

14

FG

F5.2E.EPS

Contact input 1

Contact output 2

Contact output 1

Analog output
4-20 mA DC

100 to 240 V AC,
50 or 60 Hz

Digital output

Contact input 2

The protective grounding for the analyzer shall be connected either the protective
ground terminal in the equipment or the ground terminal on the case.
Standard regarding grounding: Ground to earth, ground resistance: 100V or less.

Figure 5.2 Wiring Connection