2 output range setting, Checking calibration contact output -18 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 86

Advertising
background image

IM 11M12A01-05E

7-13

7. Startup

7.7.2

Output Range Setting

This section describes how to set the analog output range.

(1) To provide an oxygen concentration, use parameter code C11 to set the minimum

oxygen concentration at 4 mA, and use parameter code C12 to set the maximum
oxygen concentration at 20 mA.

(2) To provide a humidity output, use parameter code C13 to set the minimum humidity

at 4 mA, and use parameter code C14 to set the maximum humidity at 20 mA.

(3) To provide a mix ratio, use parameter code C15 to set the minimum mix ratio at 4

mA, and use parameter code C14 to set the maximum mixing ratio at 20 mA.

Refer to Table 7.6 for the parameter codes. For more details, consult Section 8.1,“Cur-
rent Output Settings,” later in this manual.

Table 7.6 Parameter codes for analog output range

T7.6.EPS

Parameter code

Set value

C01

0 Oxygen concentration

1 Humidity

2 Mix ratio

C11 Minimum oxygen concentration (at 4 mA)
C12 Maximum oxygen concentration (at 20 mA)
C13 Minimum humidity (at 4 mA)
C14 Maximum humidity (at 20 mA)
C15 Minimum mix ratio (at 4 mA)
C16 Maximum mix ratio (at 20 mA)

Advertising