5 wiring for contact input, 1 cable specifications, 2 wiring procedure – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 70: Zr202g high-temperature humidity analyzer -1

Advertising
background image

IM 11M12A01-05E

5-7

5. Wiring

5.5 Wiring for Contact Input

The converter can execute specified function when receiving contact signals.

To use these contact signals, proceed wiring as follows:

Converter

Terminal box

DI-1

DI-C

DI-2

Contact input 1

Contact input 2

F5.7E.EPS

Figure 5.14 Contact Input Wiring

5.5.1

Cable Specifications

Use a 2-wire or 3-wire cable for this wiring. Depending on the number of input(s),
determine which cable to use.

5.5.2

Wiring Procedure

(1) M4 screws are used for the terminals of the converter. Each wire in the cable should

be terminated in the corresponding crimp-on terminal.

(2) The ON/OFF level of this contact input is identified by the resistance. Connect a

contact input that satisfies the descriptions in Table 5.2.

Table 5.2 Identification of Contact Input ON/OFF

T5.2E.EPS

Closed

Open

Resistance

200 V or less

100 kV or more