4. Piping

4.3 Piping for System Configuration 3

Piping in System 3 is illustrated in Figure 4.9. In System 3, calibration is automated;
however, the piping is basically the same as that of System 2. Refer to Section 4.2.

Adjust secondary pressure of both the air set and the zero gas regulator so that these two
pressures are approximately the same. The flow rate of zero and span gases (normally
instrument air) are set by individual needle valve. After installation and wiring, check
the calibration contact output (see Sec. 7.9.2), and adjust zero gas regulator and calibra-
tion gas needle valve so that zero gas flow is within the permitted range. Next check
span gas calibration contact output and adjust air set so that span gas flow is within the
permitted range.

~

100 to 240 V AC

Auto Calibration unit

ZR20H

Reference gas

Calibration gas

Contact input
Analog output, contact output
Digital output (HART)

Air Set

Instrument air

Calibration gas
unit case

Calibration gas pressure regulator

Zero gas cylinder

F4.9E.EPS

Model ZR202G Integrated type Zirconia
High Temperature Humiddity Analyzer

Span gas

Figure 4.9 Piping for System 3