Preference Order of Output Hold Values

The output hold value takes the following preference order:

During error occurrence

During calibration

During maintenance

During warm-up

Preference order (high)

8.2.2E.siki

For example, if the output current is set to 4 mA for “under maintenance”, and no
output-hold output for during calibration is preset, the output is held at 4 mA during the
“ under maintenance” display. However, the output hold is released at the time of
starting the calibration, and the output will be again held at 4 mA after completing the
calibration and when the output stabilization time elapses.

8.2.3

Output Hold Setting

Table 8.4 lists parameter codes with set values for individual set items.

Table 8.4 Parameter Codes for Output Holding

Set items

Parameter code

Set value

During warm-up

C04

0: 4 mA
1: 20 mA
2: Holds Set value

During maintenance C05

0: Without hold feature
1: Last measured value..
2: Holds set values.

During calibration

C06

0: Without hold feature
1: Last measured value.
2: Holds set values.

During error occurrence C07

0: Without hold feature
1: Last measured value.
2: Holds set values.

T8.5E.EPS

8.2.4

Default Values

When the analyzer is delivered or data are initialized, the output current settings are by
default as shown in Table 8.5.

Table 8.5 Output Current Default Values

Status

Output hold (min. and max. values)

During warm-up

4 mA

Under maintenance

Holds output at value just before maintenance started.

Under calibration or blow-back

Holds output at value just before starting calibration

or "blow-back."

On Error occurrence

Holds output at a preset value.

T8.6E.EPS

Preset value

4 mA

4 mA

4 mA

3.4 mA