5.4 Wiring for Contact Output

The equipment can output a maximum of two contact signals. These contact outputs can
be used for different applications such as a low-limit alarm or high-limit alarm.

Do the contact output wiring according to the following requirements.

DO-1
DO-1

DO-2
DO-2

F5.6E.EPS

#2 Output

Analyzer

Terminal box

Annunciator or the like

#1 Output

Figure 5.6 Contact Output Wiring

5.4.1

Cable Specifications

The number of wires in cable varies depending on the number of contacts used.

5.4.2

Wiring Procedure

(1) M4 screws are used for the terminals. Use crimp-on terminals appropriate for M4

terminal screws for cable connections.

(2) The contact output relays are rated 30 V DC 3 A, 250 V AC 3 A. Connect a load

(e.g. pilot lamp and annunciator) within these limits.