Calibration, 1 calibration briefs, Calibration gas -3 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 118: Compensation -4

Advertising
background image

IM 11M12A01-05E

9-1

9. Calibration

9. Calibration

The following describes the calibration procedures for the ZR202G Zirconia High-
temperature Humidity Analyzer (integrated model).

9.1 Calibration Briefs

9.1.1

Measurement Principle of Zirconia Humidity Analyzer

A solid electrolyte such as zirconia allows the conduction of oxygen ions at high
temperatures. Therefore, when a zirconia-plated element with platinum electrodes on
both sides is heated up in contact with gases having different partial-oxygen pressures on
each side, oxygen ions flow from a high partial-oxygen pressure to a low partial-oxygen
pressure, causing a voltage. When a sample gas introduced into the zirconia-plated
element with the measurement electrode, and air (21.0 vol% O

2

) is flowed through the

reference electrode, an electromotive force (mV) is produced between the two elec-
trodes, governed by Nernst’s equation as follows:

E = - (RT/nF) log

e

(y/a) ………………………… Equation (1)

where, R = Gas constant

T = Absolute temperature
n : 4

F = Faraday’s constant

y = O

2

vol% on the zirconia element measurement electrode

a = O

2

vol% to 21.0 vol % O

2

on the zirconia element reference electrode

The humidity analyzer uses a sample gas composed of water vapor and air.

(A) For the vol% H

2

O measurement

x : Assuming that H

2

O vol% in a mixed gas is measured:

y = (100 – x) 3 0.21 …………………. Equation (2)
From the above equations (1) and (2), we obtain:

E = -K log y/a = -Klog [(100 – x) 30.21 /21]
= - K log (1 –0.01 x) ……………… Equation (3)
where, K = Constant
Using the above equation (3), we can calculate the water vapor in vol% from the
electromotive force.

Advertising