2 setting output contact, Input contact settings -18, Warning – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 111

Advertising
background image

IM 11M12A01-05E

8-14

8.4.2

Setting Output Contact

Set the output contacts following Table 8.9.

Table 8.9 Parameter Codes for Output Contact Setting

Set item

Parameter

Set value

code

Output contact 1

Operation

E10

0 Operated in closed status. (Normally deenergized)
1 Operated when open. (Normally energized) (Note 1)

Error

E20

0 Not operated if an error occurs.
1 Operated if an error occurs.

High-high limit alarm

E21

0 Not operated if a high-high limit alarm occurs.
1 Operated if a high-high limit alarm occurs. (Note 2)

High-limit alarm

E22

0 Not operated if a high-limit alarm occurs.
1 Operated if a high-limit alarm occurs. (Note 2)

Low-limit alarm

E23

0 Not operated if a low-limit alarm occurs.
1 Operated if a low-limit alarm occurs.

Low-low limit alarm

E24

0 Not operated if a low-low limit alarm occurs.
1 Operated if a low-low limit alarm occurs. (Note 2)

Maintenance

E25

0 Not operated during maintenance.
1 Operated during maintenance (see Section 8.2.1).

Calibration

E26

0 Not operated during calibration.
1 Operated during calibration (see Section 8.2.1).

Range change

E27

0 Not operated when changing ranges.
1 Operated when changing ranges. (Note 3)

Warm-up

E28

0 Not operated during warming up.
1 Operated during warming up.

Calibration-gas

E29

0 Not operated while a calibration-gas pressure drop contact is being closed.

pressure drop

1 Operated while a calibration-gas pressure drop contact is being closed. (Note 4)

Unburnt gas

E32

0 Not operated while a unburnt gas detection contact is being closed.

detection

1 Operated while a unburnt gas detection contact is being closed. (Note 5)

T8.10E.EPS

Note 1: Output contact 2 remains closed.
Note 2: For the high-high alarm, the concentration alarm must be preset (see

Section 8.3).

Note 3: Range change answer-back signal. For this action, the range change must be

preset during the setting of input contacts (Section 8.5).

Note 4: Calibration gas pressure decrease answer-back signal. Calibration gas pressure

decrease must be selected beforehand during the setting of input contacts.

Note 5: Non-combusted gas detection answer-back signals. “Non-combusted gas”

detection must be selected during the setting of input contacts.

WARNING

• Output contact 2 is linked to the detector’s heater power safety switch. As such, if

output contact 2 is on, the heater power stops and an Error 1 (cell voltage abnormal)
or Error 2 (heater temperature abnormal) occurs.