IM 11M12A01-05E

8-16

8.5 Input Contact Settings

The equipment input contacts execute set functions by accepting a remote contact signal.
Table 8.11 shows the functions executed by a remote contact signal.

Table 8.11 Input Contact Functions

T8.5.0.1.EPS

Set item

Function

Calibration-gas pressure

While the contact signal is on, neither semi-automatic nor

decreased

automatic calibration is possible.

Measuring range change

While contact input is On, range of Analog Output is switched
as follows:
When analog output range is set to "Humidity", then output range is
switched to 0 to 100% H

2

O.

When analog output range is set to "Mixing ratio", then output range
is switched to 0 to 1 kg/kg.
When analog output range is set to "Oxygen", then range is switched
to 0 to 25% O

2

.

Calibration start

If the contact signal is applied, semi-automatic calibration starts
(only if the semi-automatic or automatic mode has been setup).
Calibration is started with an applied one-second time interval
single-output contact signal. Even if a continuous contact
signal is applied, a calibration is not repeated. If you
want to perform a second calibration, turn the contact signal off
and then back on.

Unburnt gas detection

If the contact signal is on, the heater power will be switched off.
(A one-second time interval single-output signal is available as
a contact signal.) If this operation starts, the sensor
temperature decreases and an error occurs. To restore it to
normal, turn the power off and then back on, or reset the
analyzer.

CAUTION

To conduct a semi-automatic calibration, be sure to set Calibration setup mode to
Semi-Auto” or “Auto.”

8.5.1

Setting Input Contact

To set the input contacts, follow the parameter codes given in Table 8.12.

Table 8.12 Parameter Codes for Input Contact Settings

Set item

Parameter code

Set value

Input contact 1 (function)

E01

0: Invalid
1: Calibration gas-pressure decrease
2: Measuring range change
3: Calibration
4: Unburnt gas detection

Input contact 2 (function)

E02

0: Invalid
1: Calibration gas-pressure decrease
2: Measuring range change
3: Calibration
4: Unburnt gas detection

Input contact 1 (action)

E03

0: Operated when closed
1: Operated when open

Input contact 2 (action)

E04

0: Operated when closed
1: Operated when open

T8.13E.EPS