9.1.3

Compensation

The deviation of a measured value from the theoretical cell electromotive force is
checked by the method in Figure 9.2 or 9.3.
Figure 9.2 shows a two-point calibration using two gases: zero and span. Cell electromo-
tive forces for a span gas with an oxygen concentration p1 and a zero gas with an
oxygen concentration p2 are measured while determining the calibration curve passing
between these two points. The oxygen concentration of the measurement gas is deter-
mined from this calibration curve. In addition, the calibration curve corrected by
calibration is compared with the theoretical calibration curve for determining the zero-
point correction ratio represented by B/A3100 (%) on the basis of A, B and C shown in
Figure 9.2 and a span correction ratio of C/A3100 (%). If the zero-point correction ratio
exceeds the range of 100630 % or the span correction ratio becomes larger than 0618
%, calibration of the sensor becomes impossible.

21.0

p1

p2

0.51

81.92

ez

e2

e1

es

0

C

A

B

Zero-point correction factor = (B/A) x 100 (%) Correctable range: 100 – 30%
Span correction factor = (C/A) x 100 (%) Correctable range: 0 – 18%

Cell
electromotive
force, mV

Calibration curve
before correction

Corrected calibration
curve (theoretical
calibration curve)

Zero origin

Span-gas concentration

Zero-gas concentration

Span origin

Oxygen concentration (vol%O

2

)

F9.2E.EPS

Figure 9.2 Calculation of a Two-point Calibration Curve and Correction Factors

using Zero and Span Gases

Figure 9.3 shows a one-point calibration using only a span gas. In this case, only the cell
electromotive force for a span gas with oxygen concentration p1 is measured. The cell
electromotive force for the zero gas is carried over from a previous measurement to
obtain the calibration curve. The principle of calibration using only a span gas also
applies to the one-point calibration method using a zero gas only.

21.0

p1

0.51

81.92

ez

e1

es

0

C

A

B

Zero-point correction factor = (B/A) x 100 (%) Correctable range: 100 – 30%
Span correction factor = (C/A) x 100 (%) Correctable range: 0 – 18%

Cell
electromotive
force, mV

Calibration curve before
correction

Previous
zero-gas data

Corrected calibration curve
(theoretical calibration curve)

Zero origin

Span origin

Span-gas concentration

Oxygen concentration (vol%O

2

)

F9.3E.EPS

Figure 9.3 Calculation of a One-point Calibration Curve and Correction Factors

using a Span Gas