4 setting purging – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 117

Advertising
background image

IM 11M12A01-05E

8-20

8.6.4

Setting Purging

Purging is to remove condensed water in the calibration gas pipe by supplying a span
calibration gas for a given length of time before warm-up of the detector. This prevents
cell breakage during calibration due to condensed water in the pipe.

Open the solenoid valve for the automatic calibration span gas during purging and after
the purge time has elapsed, close the valve to start warm-up.

Purging is enabled when the cell temperature is 1008 C or below upon power up and the
purge time is set in the range of 1 to 60 minutes.

F8.03.EPS

Displayed alternately

Figure 8.3 Display during Purging

8.6.4.1 Procedure

Use the parameter-code table below to set the purging time.

The allowable input ranges from 0 to 60 minutes.

Table 8.17 Parameter Code

Set item

Parameter code Set range Units

T8.17E.EPS

Purging time

F15

0 to 60

minutes

8.6.4.2 Default Value

When the analyzer is delivered, or if data are initialized, purging time is set to 0
minutes.