2 calibration procedures, 1 calibration setting, Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 123

Advertising
background image

IM 11M12A01-05E

9-6

9.2 Calibration Procedures

CAUTION

Calibration should be made under normal operating conditions (if the probe is connected
to a dryer, the analyzer will undergo calibration under the operating conditions of the
dryer). To make a precise calibration, conduct both zero-point and span calibrations.

9.2.1

Calibration Setting

The following sets forth the required calibration settings:

9.2.1.1Mode

There are three calibration modes available:
(1) Manual calibration which allows zero and span calibrations or either one manually in

turn;

(2) Semi-automatic calibration which lets calibration start with the touchpanel or a

contact input, and undergoes a series of calibration operations following preset
calibration periods and stabilization time; and

(3) Automatic calibration which is carried out automatically following preset calibration

periods.

Calibrations are limited by the following mode selection:

• When manual calibration is selected:

Manual calibration only can be conducted. (This mode does not allow semi-automatic
calibration with a contact input nor automatic calibration even when its start-up time
has reached.)

• When semi-automatic calibration is selected:

This mode enables manual and semi-automatic calibrations to be conducted.
(The mode, however, does not allow automatic calibration even when its start-up time
has reached.)

• When automatic calibration is selected:

This calibration can be conducted in any mode.

9.2.1.2Calibration Procedure

Select both span and zero calibrations or span calibration only or zero calibration only.
Usually select span and zero calibrations.

9.2.1.3Zero-gas Concentration

Set the oxygen concentration for zero-point calibration. Enter the oxygen concentration
for the zero gas in the cylinder used.

9.2.1.4Span-gas Concentration

Set the oxygen concentration for span calibration. If instrument air is used as the span
gas, enter 21 %O

2

.

When using the ZO21S Standard Gas Unit (for use of the atmospheric air as a span gas),
use a hand-held oxygen analyzer to measure the actual oxygen concentration, and then
enter it.