<Treatment after completion of calibration>
(1) Fully close the needle valve mounted on the calibration gas inlet of the detector.
(2) Remove the tube connecting the detector to the standard gas unit.

Store the standard gas unit with the gas cylinder mounted where the ambient tempera-
ture does not exceed 408 C. Otherwise, the gas cylinder may explode. Store the spare
gas cylinders under the same condition.

WARNING