Initial settings should be set to match the plant conditions when installing the
converter. Current output data settings, alarm data settings, contact data settings,
other settings.

Display and setting content:
Display Related Items : Oxygen concentration (vol% O

2

), Moisture quantity (vol%

H

2

O), mixture ratio (kg/kg), relative humidity (%RH), dew point (8 C), cell tempera-

ture (8 C), thermocouple reference junction temperature (8 C), maximum/minimum/
average oxygen concentration (vol% O

2

), maximum/minimum/average moisture

quantity (vol% H

2

O), maximum/minimum/average mixture ratio (kg/kg), cell e.m.f.

(mV), output 1, 2 current (mA), cell response time (seconds), cell internal resistance
(V), cell condition (in four grades), heater on-time rate (%), calibration record (ten
times), time (year/month/day/hour/minute)

Calibration Setting Items : Span gas concentration (vol% O

2

), zero-gas concentration

(vol% O

2

), calibration mode (auto, semi-auto, manual), calibration type and method

(zero-span calibration, zero calibration only, span calibration only), stabilization
time (min.sec), calibration time (min.sec), calibration period (day/hour), starting
time (year/month/day/hour/minute)

Output Related Items : Analog output/output mode selection, output conditions when

warming-up/maintenance/calibrating/abnormal, oxygen concentration at 4mA/
20mA (vol% O

2

), moisture quantity at 4mA/ 20mA (vol% H

2

O), mixture ratio at

4mA/ 20mA (kg/kg), time constant, preset values when warming-up/maintenance/
calibrating/abnormal, output preset values on abnormal

Alarm Related Items : Oxygen concentration high-alarm/high-high alarm limit values

(vol% O

2

), Oxygen concentration low-alarm/low-low alarm limit values (vol% O

2

),

Moisture quantity high-alarm/high-high alarm limit values (vol% H

2

O), moisture

quantity low-alarm/low-low alarm limit values (vol% H

2

O), mixture ratio high-

alarm/high-high alarm limit values (kg/kg), mixture ratio low-alarm/low-low alarm
limit values (kg/kg), oxygen concentration alarm hysteresis (vol% O

2

), moisture

quantity alarm hysteresis (vol% H

2

O), mixture ratio alarm hysteresis (kg/kg),

oxygen concentration/moisture quantity/ mixture ratio detection, alarm delay
(seconds)

Contact Related Items : Selection of contact input 1 and 2, selection of contact output

1 and 2 (abnormal, high-high alarm, high-alarm, low-alarm, low-low alarm,
maintenance, calibrating, range switching, warming-up, calibration-gas pressure
decrease, flameout gas detection

Converter Output : One mA analog output point (4 to 20 mA DC (maximum load

resistance of 550 V)) with mA digital output point (HART) (minimum load
resistance of 250 V).
Range: any setting between 0 to 25 through 0 to 100 vol% H

2

O, and partial range is

available (Maximum range value/minimum range value 1.3 or more)
For the log output, the minimum range values are fixed to 0.1 vol% O

2

for the

oxygen concentration, 0.1 vol%H

2

O for the moisture quantity, and 0.01 kg/kg for

the mixture ratio.
4 to 20 mA DC linear or log can be selected.
Input/output isolation
Output damping : 0 to 255 seconds.
Hold/non-hold selection, preset value setting possible with hold.

Contact Output : Two points, contact capacity 30V DC 3A, 250V AC 3A (resistive

load)
Normally energized or normally de-energized can be selected.
Delayed functions (0 to 255 seconds) and hysteresis function (0 to 9.9 vol% O

2

) can

be added to high/low-alarms.
The following functions are programmable for contact outputs.