4 other equipment, 1 stop valve (part number: l9852cb or g7016xh), 2 check valve (part number: k9292dn or k9292ds) – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 34: Other equipment -16, Stop valve (part number: l9852cb or g7016xh) -16, Check valve (part number: k9292dn or k9292ds) -16

Advertising
background image

IM 11M12A01-05E

2-16

2.4 Other Equipment

2.4.1 Stop Valve (part number: L9852CB or G7016XH)

This valve is mounted on the calibration gas line in the system to allow for one-touch
calibration. This applies to the system configuration shown for System 1 in section1.

Standard Specifications
Connection : Rc 1/4 or 1/4 FNPT

Material : SUS 316 (JIS)

Weight : Approx. 80 g

Joint: RC 1/4, Material: SUS 316 (JIS)

Joint: 1/4 NPT, Material: SUS 316 (JIS)

T2.9E.EPS

Description

Part No.

L9852CB

G7016XH

F15.EPS

(Full open length)

40

Rc1/4 or 1/4NPT

43

55

2.4.2 Check Valve (part number: K9292DN or K9292DS)

This valve is mounted on the calibration gas line (directly connected to the detector).
This is applied to a system based on the system configuration (System 2 and 3).

This valve prevents the process gas from entering the calibration gas line. Although it
functions as the stop valve, operation is easier as it does not require opening/closing at
each calibration.

Screw the check valve into the calibration gas inlet of the detector instead of the stop
valve.

Standard Specification
Connection : Rc1/4 or 1/4FNPT

Material : SUS304 (JIS)

Pressure : 70 kPa G or more and 350 kPa G or less

Weight: Approx. 40 g

Joint: RC 1/4, Material: SUS304 (JIS)

Joint: 1/4 NPT, Material: SUS304 (JIS)

T2.10E.EPS

Part No.

K9292DN

K9292DS

Description