Equipment setup and others in group f – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 144

Advertising
background image

IM 11M12A01-05E

10-17

10. Other Functions

Equipment Setup and Others in Group F

T10.6F.EPS

Code Item

Tuning

Engineering Default setting

unit

F01 Equipment setup

0: Oxygen analyzer

Not initialized

1: Humidity analyzer

F02

F04 Selection of temperature units

0: degree C

degree C

1: degree F

F05 Selection of pressure units

0: kPa

kPa

1: psi

F08 Selection of display items

0: Oxygen concentration

Humidity

1: Humidity

2: Mixed ratio

3: Item selected with
analog output

F10 Date

YY.MM.DD/H

H.MM

F11 Period over which average

1 to 255 hours

Hours

One hour

values are calculated

F12 Period over which max. and

1 to 255 hours

Hours

24 hours

min. values are monitored

F13 Process gas temperature

0 to 3000

8C

300 8C

F14 Process gas pressure

0 to 300

kPa abs. 101.33 kPa abs.

F20

F21

F22

F23

F30 Initializing all data

F31 Initializing data in group A

F32 Initializing data in group B

F33 Initializing data in group C

F34 Initializing data in group D

F35 Initializing data in group E

F36 Initializing data in group F

Advertising