3 - 5

IM 11G02M01-01E

(7) Example of sampling system configuration

The system configuration may vary depending upon the nature of measured gas, coexistent gases
or application. A typical configuration diagram is shown in Fig. 3-2. Since a system configura-
tion depends upon measured gas, consult with Yokogawa.

Fig. 3-2 A typical example of sampling system

Pump

4-way selector
cock

Primary filter

Safety
drain
trap

Electronic
dehumidifier

Drain
pot

Standard gas

for zero calibration

Standard gas for
span calibration

Ball valve

Drain

Sample
gas inlet

Flow meter

Exhaust
(atmospheric
pressure)

Secondary
filter

Gas

analyzer

(IR200)

External

O

2

analyzer

F03.EPS

1

6

2

3

5

4

7

8

9

10

Measurement of sample gas with low moisture content (room-temperature saturation level or below):
CO, CO

2

and

CH

4

measurement

ء

ء

A dehumidifier must be used for CO (0-200 ppm)
and SO

2

measurements (to prevent saturation at

2

Њ

C). Even in any other gas measurements, if

sample gas has a high moisture content or if
condensation is likely to occur, a safety drain trap
and a dehumidifier should be used.

ء

Pump

Primary filter

Safety
dran trap

Electronic

dehumidifier

Drain

pot

Standard gas

for span calibration

Ball valve

Drain

Flow meter

Secondary filter

Gas analyzer

(IR200)

External O

2

analyzer

Measurement of sample gas with high moisture content or NO, SO

2

, or CO (0-200 ppm range) measurement

A dehumidifier must be used for NO, SO

2

, and CO (0-200 ppm) measurements (to prevent

saturation at 2

Њ

C). Use either atmospheric air or cylinder gas as a zero calibration gas and

supply it through a bubbler (humidifying) to reduce interference of water.

Atmosphere

NO

2

/NO

converter

Pump

ء

ء

3-way

solenoid valve

Solenoid

valve

Solenoid

valve

Bubbler

Exhaust
(atmospheric pressure)

Solenoid

valve

Standard gas

for zero calibration

p

1

2

3

3

4

5

6

8

8

9

14

14

14

10

12

14

11

13

No.

Item

Description

Primary filter (mist filter)

Remove dust and mist

Safety drain trap

Pump

Electronic dehumidifier

Drain pot

Ball valve

4-way selector cock

Standard gas for calibration

Flow meter

Separate and discharges drain

Suck in sample ga

Dehumidify sample gas

Collect discharged water from dehumidifier

Used for discharging drain

Used for switching sampling and calibration lines

Used for zero/span calibration

Adjust and monitor sample gas flow rate

Secondary filter (membrane filter)

Remov fine dust

T03.EPS

NO

2

/NO converter

Convert NO

2

gas into NO gas

3-way solenoid valve

Used for introducing humidified gas

Bubbler

Humidify calibration gas

Solenoid valve

Used for switching sampling and calibration lines

1

2

3

4

5

6

7

8

9

10

11

12

13

14