Yokogawa IR200 User Manual

Page 77

Advertising
background image

7 - 8

IM 11G02M01-01E

Fig. 7-3 Configuration of measuring unit (2-component analyzer: block cell + pipe cell)

16

15

11

13

12

14

12

13

11

10

3

10

17

9

1

7

8

6

5

2

5

4

18

No.

Name

1

Screw (for fixing light source unit)

2

Screw (for fixing detector)

3

Base plate

4

Light source unit

5

Screw (for fixing block cell)

6

Block cell

7

Infrared transmission window

8

O-ring

9

Detector

10

Screw (for fixing base plate)

11

Support

12

Screw (Fixing cell retainer)

13

Cell retainer

14

Pipe cell

15

O-ring

16

Infrared transmission window

17

Screw (fixing detector)

18

Detector

3) How to remove measuring unit (See Fig. 7-3)

1. For Step 1. to 4., see Item 7.5.1(1), How to remove pipe cell.5. Remove the detector output

cord connector from the printed board.

6. Remove wiring to the 2-pin terminals of the infrared ray light source assembly and chopper

motor pin connector (No. 8) from the printed board.

7. Detach the 4 screws (No. 3) fastening the base plate (No. 4) to remove the measuring unit.

Note) Special care should be taken when assembling or disassembling the measuring cell

to avoid the application of force to the detector pipe or infrared ray light source unit
pipe. If the pipe is deformed or damaged by excessive force, there is a danger of gas
leak, thus resulting in misoperation.