2 forced stop of auto zero calibration – Yokogawa IR200 User Manual

Page 56

Advertising
background image

6 - 21

IM 11G02M01-01E

6.5.2 Forced stop of auto zero calibration

This mode is used to cancel the auto zero calibration forcedly.

1. In the User mode that is displayed, point

the cursor to “Setting of Auto Zero
Calibration” by pressing the

or

key. Press the

ENT

key.

2. In the “Setting of Auto Zero Calibration”

item selection screen that appears, point
the cursor to “Stop Auto Zero Calibra-
tion” by pressing the

or

key.

Press the key.

3. “Stop Auto Zero Calibration” is inverted.

A message appears, prompting you to
verify that you want to cancel or continue
auto zero calibration. To cancel the auto
zero calibration forcibly, press the

ENT

key. If you press the

ESC

key, auto zero

calibration is cancel (not stopped).

(

)

ENT

(

)

ENT