6.2 Calibration setting

This mode is used to set calibration concentration and actions. The calibration setting involves
calibration concentration, calibration gas, zero calibration, calibration range and auto-calibration
component.

“Calibration Gas” is displayed if NO or SO

2

is included in components to be analyzed.

6.2.1 Setting of calibration concentration

It allows you to set concentrations of the standard gas (zero and span) of each channel used for
calibration.

1. During measurement, press the

MODE

key

to display the User mode.

2. Point the cursor to “Setting about Cali-

bration” by pressing the

or

key.

Press the

ENT

key.

3. In the “Setting about Calibration” screen

that appears, point the cursor to “Calibra-
tion Value” by pressing the

or

key. Press the

ENT

key.

4. In the “Calibration Concentration CH

Selection” screen that appears, point the
cursor to CH you want to set by using the

or

key. Press the

ENT

key.