6 - 11

IM 11G02M01-01E

5. In the “Auto Calibration component

Selection” screen that appears, select
“enable” or “disable” by pressing the
or

key. After setting, press the

ENT

key.

To close “Setting of Auto Calibration

Component” or to cancel this mode midway,

press the

ESC

key.

A previous screen will return.

To close Auto Calibration
Component setting

ESC

(

)

ENT

End of Auto Calibtation Component Setting

The components which were set to enable is calibrated to zero at once at the time of

auto calibration regardless of setting in 6.2.3 Setting of manual zero calibration .

Caution

Auto calibration is made in the following rules according to the setting.
1. Zero calibration at once of CHs (components) which were set to enable .
2. The span of CHs (components) which were set to enable is calibrated in the ascending
order of CH number.

Example 1. In case all of CH1: CO

2

, CH2: CO, CH3: O

2

, were set to enable .

Zero calibration (at once) of CH1 to CH3 span calibration of CH1 (CO

2

)

span calibration of CH2 (CO) span calibration of CH3 (O

2

)

Example 2. In case, out of CH1: CO

2

, CH2: CO and CH3: O

2

, CH1

(

CO

2

) was set to enable ,

CH2 (CO) was set to enable an CH3 (O

2

) was set to disable .

Zero calibration (at once) of CH1 and 2 span calibration of CH1 (CO

2

)

span calibration of CH2 (CO)

Example