3 setting of manual zero calibration – Yokogawa IR200 User Manual

Page 41

Advertising
background image

6 - 6

IM 11G02M01-01E

1. During measurement, press the

MODE

key

to display the User mode.

2. Point the cursor to “Setting about Cali-

bration” by pressing the

or

key.

Press the

ENT

key.

3. In the “Setting about Calibration” screen

that appears, point the cursor to “About
ZERO Calibration” by pressing the

or

key. Press the

ENT

key.

4. In the “CH Selection” screen that

appears, point the cursor to CH you want

to set by using the

or

key. Press

the

ENT

key.

(

)

ENT

(

)

ENT

MODE

(

)

ENT

6.2.3 Setting of manual zero calibration

If zero calibration is to be made manually, select whether to calibrate all components at once or
each of them separately upon selection.