6 - 28

IM 11G02M01-01E

Average period

It allows you to set an average period of the
average value of O

2

correction.

It enables you to set an average time of 1 to 59
minutes (1-minute step) or 1 to 4 hours (1-hour
step).
Changing the setting resets the average value of
O

2

correction and O

2

average value. (Pressing

the ENT validates the resetting only for compo-
nents whose setting was changed.)

Maintenance mode

To open the maintenance mode, enter a password. After entering the password, press the

ENT

key.

The password can be used for the Password Setting in the Maintenance mode. A password is set
to “0000” before factory-shipment. This value is available for the Maintenance mode.

Suppose the average period is 1 hour.

• Sampling occurs every 30 seconds.

• Every 30 seconds, the average for last 1 hour (time setting) is output.

• At the instant of resetting, zero is assumed for all past values. It means that the average value

will not be correct for 1 hour after resetting.

Example of average action

Average value

Time

Reset

Time