4 setting of auto calibration, 1 auto calibration – Yokogawa IR200 User Manual

Page 50

Advertising
background image

6 - 15

IM 11G02M01-01E

6.4 Setting of auto calibration

6.4.1 Auto calibration

Auto calibration is automatically carried out
when zero calibration and span calibration are
set.

Before changing the setting of auto calibration,
set the ON/OFF to “OFF”.

1. During measurement, press the

MODE

key

to display the User mode.

2. Point the cursor to “Setting of Auto

Calibration” by pressing the

or

key. Press the

ENT

key.

3. In the “Setting of Auto Calibration”

screen that appears, point the cursor to
any item you want by pressing the

or

key. Press the

ENT

key.

4. In the “Auto Calibration Parameter

Setting” screen that appears, perform the
value entry or the setting. For the value
entry or setting change, use the

or

key. To change the setting, use the
key to move the cursor to the right.

After setting, press the

ENT

key, and auto

calibration is carried out by the entered
setting value.

(

)

ENT

MODE

(

)

ENT

ENT

End of Auto Calibration Setting

Description of setting items

• Start Time : Setting at the first calibration

(day of the week, hour, minute)

• Cycle

: A period between the start time of one

calibration and another
(unit : hour/day)

• Flow Time : The time required for the calibration gas

to be replaced in the cell

• ON/OFF : Auto zero calibration ON or OFF

To close the "Setting of Auto calibration" or cancel

this mode midway, press the

ESC

key.

A previous screen will return.

To close Setting of Auto calibration

ESC