Yokogawa IR200 User Manual

Page 65

Advertising
background image

6 - 30

IM 11G02M01-01E

It enables you to set a password to be used when
switching the parameter setting mode to the
maintenance mode.

Description of Password Setting screen

Note) The password set herein must be

managed for safety. Failure to enter
the correct password will not open the
Maintenance mode.

• O

2

correction reference value setting screen

• Password Setting screen

Description of O

2

corection

reference value setting screen

O

2

correction value is calculated from the

following equation by values set herein:

O

2

correction concentration value

21 set value

21 - O

2

concentration

measured value

component
concentration
measured value

=

The value of the fractional part in this calculation
is limited to 4.
If it is 4 or more, it is fixed at 4.

x

ؒ

O

2

correction reference value

: 00 to 19% in 1% step

(initial value: 12%)

Setting range

• Station No. setting screen (option)

Setting range

• Station No. : 00 to 31 (initial No.: 00)

* Please refer to another manual (IM 11G02P01-01E)

about the communication function.

Advertising