Yokogawa IR200 User Manual

Page 27

Advertising
background image

3 - 14

IM 11G02M01-01E

(9) Timing of calibration contact output

1. Manual calibration (See “6.9 Calibration”).

Pump ON/OFF contact

• Zero calibration

Zero calibration contact

Output signal hold

Hold extension time

Calibration end

Output hold function

(with hold ON setting)

CH1 to CH4 span calibration
contact

on

off
on

off

off

Pump ON/OFF contact

• Span calibration

Zero calibration contact

Output signal hold

Hold extension time

Note)

Calibration end

Span 1 calibration contact

Span 2 calibration contact

Output hold function

(with hold ON setting)

off

off

off

on

on

off

The hold extension time
depends on the gas
flow time of the
automatic calibration
settings.

SPAN

ZERO

ENT

ENT

ENT

ENT

When selecting

and keys.

CH2 using

2. In case of automatic calibration (example shown in 6.4.1, Automatic calibration settings)

Automatic

calibration

start

Zero

calibration

CH2 span gas

350s

CH1

span

calibration

CH2

span

calibration

CH3 span calibration

(automatic calibration end)

Pump ON/OFF contact

Zero calibration contact

CH1 span calibration contact

CH2 span calibration contact

CH3 span calibration contact

Automatic calibration contact

Output hold function

CH3 span gas

350s

Zero gas

350s

CH1 span gas

350s

Hold extension time

Output signal hold

(With hold ON setting)