6 - 5

IM 11G02M01-01E

(5) On the “Gas Selection” screen that

appears, select “with” or “without” of
moisture content in calibration gas using

the

or the

key.

• Select “with” to have the calibration gas

contain a certain amount of moisture.

• Select “without” to use dry gas (moisture

content: saturated at 70

°

C or lower) such

as cylinder gas as calibration gas.

• When “without” is selected for zero gas,

“with” cannot be selected for span gas.

After the selection is made, press the

ENT

key.

To terminate or cancel calibration gas setting,

press the key.

The previous screen appears again.

To close the setting

ESC

Note: Be sure to make the setting according to the sampling system created. We recommend

you to select “with” moisture in zero gas for the measurement of NO and SO

2

. In this

case, be sure to perform bubbling (humidification) of zero gas to make its moisture
content kept at constant level before feeding the gas. (See “(7) Example of sampling
system configuration” in “3.4 Sampling system.”)

(

)

ENT

End of Calibration gas

Setting