Average and o – Yokogawa IR200 User Manual

Page 24

Advertising
background image

3 - 11

IM 11G02M01-01E

Terminal block 3

<TN3>

11

12

13

14

15

16

17

18

19

20

1

2

3

4

5

6

7

8

9

10

(M3.5 screw)

Instrument error
(DO)

CH4 range identification
signal output (DO)

CH3 range identification
signal output (DO)

CH2 range identification
signal output (DO)

CH1 range identification
signal output (DO)

Auto calibration

remote start

input (DI)

Average value reset

input (DI)

Remote hold input

(DI)

Unused

Unused

Terminal block 3 <TN3>

Between 1 to 4

: For internal connection. Must

not be wired. (Must not be used
as junction terminal.)

Between 5 – 6

: Remote hold input. No hold

when open. Output hold when
short-circuited.

Between 7 – 8

: Average value reset input.

Short-circuiting the contact input
(for at 1.5 sec or more) resets O

2

average and O

2

correction

average simultaneously.
Opening it restarts the average
value.

Between 9 – 10 : Automatic calibration remote

start input. Open input after
strapping for at least 1.5 seconds
starts the automatic calibration
whether automatic calibration
setting is ON or OFF.

Between 11 – 12 : Conductive when analyzer unit

error is producted. Normally
open.

Between 13 – 14 : CH4 range identification signal

Between 15 – 16 : CH3 range identification signal

Between 17– 18 : CH2 range identification signal

Between 19– 20 : CH1 range identification signal

Note) Range identification

signal is conductive at
Low range or open at
High range.
In case of 1-range
system, the signal
remains open.