Setting and calibration, 1 changeover of range – Yokogawa IR200 User Manual

Page 36

Advertising
background image

6 - 1

IM 11G02M01-01E

6. SETTING AND CALIBRATION

6.1 Changeover of range

This mode is used to select the ranges of measured
components.

1. During measurement, press the

MODE

key to

display the User mode.

2. Point the cursor to “Changeover of Range”.

Press the

ENT

key.

3. The “Channel Selection” screen appears.

Press the “

“ or “

” key until the “ ”

cursor moved selects a desired CH (compo-
nent).

4. After selection, press the

ENT

key.

Note) The range of O

2

correction instanta-

neous values and O

2

correction aver-

age values is automatically switched by
changing the range of instantaneous
value of each CH (component).

5. In the Range Setting screen that appears,

move the cursor by pressing the

or

key to select the range. (The range with a
mark of is currently selected.)

6. After selection, press the

ENT

key.

7. Measurement is conducted within the selected

range. The range identification signal (CO) is
shorted with the low range (Range 1), and
open with the high range (Range 2).

Note) If the Remote Range is set to ON, the

changeover of range cannot be per-
formed on the screen.

To close Changeover of range, or cancel the

commad midway, press the key.

A previous screen will return.

To close Changeover of range

ESC

MODE

(

)

ENT

ENT

(

)

ENT

End of Range Changeover

Advertising