2 hysteresis setting – Yokogawa IR200 User Manual

Page 49

Advertising
background image

6 - 14

IM 11G02M01-01E

1. In the “Alarm Setting CH Selection”

screen that appears, point the cursor to
“Hysteresis” by pressing the

or

key. Press the

ENT

key.

2. In the “Hysteresis Value Setting” screen

that appears, enter hysteresis values.

For the value entry, 1-digit value is
increased or decreased by pressing the

or

key, and pressing the

key

moves the digit. After setting, press the

ENT

key to carry out hysteresis.

Hysteresis (In case of upper limit alarm)

If hysteresis values exceed the upper limit value as shown in graph, an alarm output is
provided. Once the alarm output is turned “ON”, it remains “ON” until the value falls
below the set lower limit of the hysteresis indication.

0 to 20% of full scale
A full scale means each range provides a full
scale of width.

Setting range

Upper limit value

ON (conductive)

OFF (open)

Hysteresis width

Alarm

(

)

ENT

End of Hysteresis Setting

Caution

The hysteresis is common to all CHs
(components).

To close the "Hysteresis Setting" or cancel

the mode midway, press the key.

A previous screen will return

.

To close Hysteresis Setting

ESC

ENT

6.3.2 Hysteresis setting

To prevent chattering of an alarm output near
the alarm setting values, set hysteresis.