External dimensions of zx8d, External connection diagram – Yokogawa IR200 User Manual

Page 95

Advertising
background image

9 - 12

IM 11G02M01-01E

Dedicated Zirconia O

2

Sensor (to be purchased

separately)
For O

2

correction, the IR200 can accept linealized 0 to 1 V

DC signal coming from an analyzer calibrated to 0 to 25%
O

2

of full scale. Dedicated zirconia O

2

sensor, Model

ZX8D, is available from Yokogawa.

Measuring method: Zirconia system

Measurable component and measuring range:

Measurable component

Minimum range Maximum range

0

Ϫ

5 vol%

0

Ϫ

25 vol%

T12.EPS

Oxygen (O

2

)

Repeatability:

Within

±

0.5% of full scale

Linearity:

Within

±

1% of full scale

Zero drift:

Within

±

1% of full scale/week

Span drift:

Within

±

2% of full scale/week

Response time: Approx. 20 seconds (for 90% response)
Sample gas flow rate: 0.5

±

0.25 L/min

Note: The Zirconia system, due to its principle, may

produce a measuring error depending on the
relative concentration versus the combustible O

2

gas concentration. Also, a corrosive gas (SO

2

of

250 ppm or more, etc.) may affect the life of the

sensor

Gas inlet/outlet size: Rc1/4

Power supply: 90 to 126 V AC or 200 to 240 V AC,

50/60 Hz

Enclosure:

Steel casing, for indoor application

Indication:

Temperature indication (LED)

Temperature alarm output:

Contact output 1a contact,
Contact capacity 220 V AC, 1 A (resistive
load)

Safety and EMC conforming standards:

Safety:

EN61010-1
Pollution degree 2 (Note)
Installation category II (Note)

Note: Installation category, called over-
voltage category, specifies impulse
withstanding voltage. Category II is for
electrical equipment.
Pollution degree indicates the degree

of existence of solid, liquid, gas or other
inclusions which may reduce dielectric
strength. Degree 2 is the normal indoor
environment.

EMC:

EN61326
EN61000-3-2
EN61000-3-3

Dimensions (H x W x D): 140

ϫ

170

ϫ

190 mm

Weight:

Approx. 3 kg

Finish color:

Munsell 5Y 7/1

ZX8D

ء

C

ء

D

Power
supply

Model Suffix code Option code

Ϫ

5

Ϫ

3

Style code

90

Ϫ

126 V AC, 50/60 Hz

200

Ϫ

240 V AC, 50/60 Hz

Dedicated zirconia O

2

sensor

Description

Style C (Non-CE conformity)
Style D (CE conformity)

T13.EPS

External Dimensions of ZX8D

Unit: mm

+

-

POWER FUSE3A

OFF

1

2

3

4

5

6

7

AC

E

OUT

ALM

800

7-M4

Gas inlet

17

Gas outlet

172.3

140

131

11

0

170

152

5

OXYGEN ANALYZER

TEMP

F08.EPS

External Connection Diagram

1

2

3

4

5

6

7

AC power
50/60Hz

supply E

Output
to IR400

Temperature
alarm output

+

-

F09-2.EPS

Caution on Safety

Before using this product, be sure to read its instruction manual in advance.

Advertising