4 setting of calibration range – Yokogawa IR200 User Manual

Page 43

Advertising
background image

6 - 8

IM 11G02M01-01E

1. During measurement, press the

MODE

key

to display the User mode.

2. Point the cursor to “Setting about Cali-

bration” by pressing the

or

key.

Press the

ENT

key.

3. In the “Setting about Calibration” screen

that appears, point the cursor to “About
Calibration Range” by pressing the

or

key. Press the

ENT

key.

4. In the “Calibration Range CH Selection”

screen that appears, point the cursor to
the CH you want to set by pressing the

or

key. Press the

ENT

key.

(

)

ENT

(

)

ENT

MODE

(

)

ENT

6.2.4 Setting of calibration range

This mode is used to set if the range of each CH (component) at the zero and span calibration
(manual calibration or auto calibration) should be calibrated with a single range or 2 ranges.