IM 11G02M01-01E

6.9 Calibration

6.9.1 Zero calibration

It is used for zero point adjustment. For zero
calibration gas, see 3.4 (3), Preparation for
standard gas in Sampling. Use a gas accord-
ing to application.

1. Press the

ZERO

key on the Measurement

screen to display the Manual Zero Cali-
bration screen.

2. Select the CH (component) to be cali-

brated by pressing the

or

key.

After selection, press the

ENT

key, and

zero gas will be supplied.

3. Wait until the indication is stabilized

with the zero gas supplied. After the
indication has been stabilized, press the

ENT

key. Zero calibration in Range

selected by the cursor is carried out.

For the CH (components) that is set to
“at once” in the “Zero Calibration” of the
Calibration Setting mode, zero calibration
is also carried out at the same time.

Caution

To close the "Zero Calibration " or

cancel this mode midway, press the

ESC

key. A previous screen will return.

To close Zero Calibration

ESC

ZERO

ENT

To Measurement screen after

executing Manual Zero Calibration

(

)

ENT