Yokogawa IR200 User Manual

Page 48

Advertising
background image

6 - 13

IM 11G02M01-01E

After setting, the alarm setting is now
completed by pressing the

ENT

key.

End of Alarm Setting

When an upper limit alarm has
occurred, the “High alarm” message
lights at CH (component)
(“Low alarm” at lower limit alarm).

Typical on-screen display when an alarm occurs

C H

O

2

0-25

vol%

C H

CO

0-100

ppm

C H

CO

2

0-10

vol%

C H

SO

2

0-100

ppm

C H

NO

X

0-100

ppm

0 0

.

0 0 0 3

.

0 0

.

2 1 0 0

.

- - - -

High alarm

Caution

For 10 minutes after turning on power, the alarm judgment is inactive.

Cursor for setting value

To close the " Alarm Setting" or to cancel

this mode midway, press the

ESC

key.

A previous screen will return.

To close the Alarm Setting

ESC

0 to 100%FS (Settable in each range)

Setting Range

ENT

Description of setting items

Upper limit value: Sets the upper limit value (concentration) of alarm.
Lower limit value: Sets the lower limit value (concentration) of alarm.
Contact action:

Selects one of upper limit alarm, lower limit alarm, and upper limit
or lower limit alarm.
Upper limit alarm ... Alarm contact operates when above upper limit value.
Lower limit alarm ... Alarm contact operates when below lower limit value.
Upper limit alarm or lower limit alarm ...Alarm contact operates when

above upper limit value or below

lower limit value.
ON/OFF: Enables the alarm function if set at "ON", or disables the alarm function if set at "OFF".

* The upper limit value cannot be set below the lower limit value, and the lower limit value

cannot be set above the upper limit value.
If it is desired to set the upper limit value below the lower limit value already stored in the
memory, reduce the lower limit value beforehand, or vice versa.