Specifications, 1 specifications, 1) standard specifications – Yokogawa IR200 User Manual

Page 84

Advertising
background image

9 - 1

IM 11G02M01-01E

9.1 Specifications

(1) Standard specifications

CO

2

, CO, CH

4

, SO

2

, NO: Non-dispersive infrared method

Single light source-single beam

O

2

:

Paramagnetic type (built-in), or
zirconia type (external)

Measurable gas components and measuring range:

• Max. 4 components measurement including O

2

.

• 1 or 2 measuring range per component.
• Measuring range ratio

1:5 (except built in O

2

)

1:20 (built in O

2

)

Max. 4 components and 2 ranges are
selectable including an O

2

measurement.

For measurable components and possible
combinations of measuring ranges, refer
to Tables 1 to 8.

Measured value indication:

Digital indication in 4 digits

(LCD with back light)

• Instantaneous value of each component
• Instantaneous value after O

2

correction

(only in CO, SO

2,

NO measurement

with O

2

)

• Average value after O

2

correction

(only in CO, SO

2,

NO measurement

with O

2

)

• O

2

average value

Analog output signals:

4 to 20 mA DC or 0 to 1 V DC, non-iso-
lated output.
Analog output corresponds to measured
value indication in 1:1.
Permissible load; 550

max. for 4 to 20

mA DC 100k

min. for

0 to 1

V DC

* Refer to Table 6, for the channel No. of

displayed values and analog output sig-
nals.

Analog input signal:

For signal input from externally installed
O

2

sensor.

Signal requirement;

(1) Signal from Yokogawa’s Zirconia O

2

sensor (Model : ZX8D*C or ZX8D*D)

(2) 0 to 1 V DC from an O

2

sensor

Input section is not isolated. This feature

is effective when an O

2

sensor is not built

in.

Relay contact output:

1a contact (250 V AC/2 A, resistive load)

Instrument error, calibration error,
range identification, auto calibration
status and maintenance status, sole-
noid valve drive for auto calibration,
pump ON/OFF.

1c contact (250 V AC/2 A, resistive load)

Upper/lower alarm contact output.
Peak count alarm contact output.

* All relay contacts are isolated mutually

and from the internal circuit.

Contact input:

Non-voltage contact (ON/0 V, OFF/5 V
DC, 5 mA flowing at ON)

Remote range changeover, auto cali-
bration remote start, remote holding,
average value resetting

Isolated from the internal circuit with a
photocoupler. Contact inputs are not iso-
lated from one another.
*Only M3.5 screw terminals are used for

all signal inputs and outputs.

Power supply: Voltage rating ;100 V AC to 240 V AC

Allowable range; 85 to 264 V AC
Frequency; 50 Hz / 60 Hz
Power consumption; 70 VA max.
Inlet

; Conform to EN60320
Protection Class 1

Operating conditions:

Ambient temperature; -5˚C to 45˚C
Ambient humidity; 90% RH max., non-

condensing

Storage conditions:

Ambient temperature; -20˚C to 60˚C
Ambient humidity; 90% RH max., non-

condensing

Dimensions (H x W x D):

19-inch rack mounting type;
177 x 483 x 493 mm

Weight:

Approx. 10 kg

Finish color:

Front panel; Off-white (Munsell 10Y7.5/

0.5 or equivalent)

Casing;

Steel-blue

Enclosure:

Steel casing, for indoor use

Material of gas-contacting parts:

Gas inlet/outlet; SUS304
Sample cell; SUS304/neoprene rubber
Infrared-ray transmitting window; CaF

2

Internal tubing; Toaron tube

Gas inlet/outlet:

Rc

1/4

or

1/4

NPT internal thread

Purge gas flow rate:

1L/min (when required)

Installation Altitude: 2000 m or less
Safety and EMC conforming standards:

Safety:

EN61010-1
Pollution degree 2 (Note)
Installation category II (Note)

Note: Installation category, called over-
voltage category, specifies impulse
withstanding voltage. Category II is for
electrical equipment.
Pollution degree indicates the degree

of existence of solid, liquid, gas or other
inclusions which may reduce dielectric
strength. Degree 2 is the normal indoor
environment.

EMC:

EN61326
EN61000-3-2
EN61000-3-3

Minimum range

Maximum range

CO

2

CO

CH

4

SO

2

O

2

(paramagnetic)

O

2

(zirconia)

0 – 200 ppm

0 – 100 vol%

0 – 100 vol%

0 – 100 vol%

0 – 5000 ppm

0 – 100 vol%

0 – 1000 ppm

0 – 500 ppm

0 – 5 vol%

0 – 500 ppm

0 – 25 vol%

0 – 5 vol%

T01.EPS

Component

Range

NO

0 – 500 ppm

0 – 5000 ppm

9. SPECIFICATIONS