3 alarm setting, 1 setting of alarm values – Yokogawa IR200 User Manual

Page 47

Advertising
background image

6 - 12

IM 11G02M01-01E

1. During measurement, press the

MODE

key

to display the User mode.

2. Point the cursor to “Alarm Setting” by

pressing the

or

key. Press the

ENT

key.

3. After the alarm setting, CH selection

screen has appeared, operate the

or

key until the cursor is aligned with a

desired CH and press the

ENT

key.

4. After the alarm item selection screen has

appeared, operate the

or

key until

the cursor is aligned with a desired item
and press the key.

(

)

ENT

MODE

(

)

ENT

Caution

Set the values so that upper limit value > lower
limit value and that (upper limit value

-

lower

limit value) > hysteresis error.
The alarm is activated in each range
independently from the settings of each range.

ENT

(

)

ENT

6.3 Alarm setting

6.3.1 Setting of alarm values

This mode is used to set the upper and lower limit value to provide an alarm output against
measured concentration during measurement. Before changing the alarm setting, set the ON/OFF
to OFF.