6 - 31

IM 11G02M01-01E

Moisture interference compensation adjustment screen

* See “7.4 Moisture interference compensation

adjustment” for details of moisture interference
compensation adjustment procedure.

Backlight Timer setting screen

Automatic OFF of the backlight in the LCD can

be selected. When the length of time set here

elapses after the measurement screen appears,

the backlight is automatically turned off.

Only when "ON" is selected, the time until

automatic OFF is displayed. Press the key

in this state, and you can change the time until

auto backlight OFF.

Press an arbitrary key while the backlight is kept

off, and the backlight comes on again.

Description of backlight timer setting screen

(

)

ENT

(

)

ENT

End of moisture interference

correction adjustment

By setting the effect of moisture content in

measured gas on this screen, the value of the

effect is subtracted to find the measurement value.

Adjustment is carried out by one of the following

methods.

• Select a desired channel (CH) and change the

• value displayed at right.

Select ALL and perform auto calculation for

all the channels in batch.

Select RESET, and the interference compensation

settings for all the channels are reset to 0.

Description of moisture interference

compensation adjustment screen