5 wiring method, Caution – Yokogawa IR200 User Manual

Page 19

Advertising
background image

3 - 6

IM 11G02M01-01E

Caution on wiring

• Wiring work must be performed with the main power set to

OFF to prevent electric shocks.

• Enforce construction of class-D grounding wire by all

means, 100

⍀ or less.

If the specified grounding construction is neglected, a
shock hazard or fault may be caused.

• Wires should be the proper one meeting the ratings of this

instrument. If using a wire which cannot endure the
ratings, a fire may occur.

• Be sure to use a power supply of correct rating. Connec-

tion of power supply of incorrect rating may cause fire.

Each external terminal is provided on the rear panel of the analyzer. (See Fig. 3-3)

Wire each terminal, referring to Fig. 3-3 and (1) to (7).

(1) Power inlet

When using supplied power cable, connect the female side to the power inlet at the rear panel of
the analyzer, and insert the male side into a receptacle matching the rating.

Fig. 3-3 Rear panel

Power inlet (3 pins)

100 to 240V AC
50/60Hz

Female
side

Male side

)

(

Supplied power cable

Fig. 3-4

Terminal block 1

Terminal block 2

Connector for
communication

Terminal block 3

Terminal block 4

Terminal block 5

3.5 Wiring method

CAUTION