Setup and initial use

1.02

Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control

103

Fig. 4-25:

Installing the safety shield

1. Hang the safety shield (1) on the edge of the heating bath (2) and position

it so that it does not interfere with the opening of the guard hood

(if present).

2. Turn eccentric screw (3) by 180° to clamp the safety shield to the heating

bath

Fig. 4-26:

Boiling temperature sensor

The boiling temperature sensor can not be used when using the temperature

sensor T auto.

1. Remove the screw cap on the condenser.
2. Insert the boiling temperature sensor in the condenser so that the white

PTFE side of the seal (2) faces the glass thread.

3. Secure the boiling temperature sensor with the screw connector cap.
4. Connect the bayonet plug (1) to the socket on the rear of the instrument

(see chapter

4.4.4, page 84).

Fig. 4-27: Temperature

sensor

Tauto

Only in conjunction with glassware sets G3, G4 and G6.

Safety shield

Boiling temperature

sensor (Laborota

4010 / 4011 and

4002 / 4003)

Temperature sensor

T auto (Laborota

4002 / 4003)