4 initial operation, 1 attaching the heating bath, 2 filling the heating bath – Heidolph LABOROTA 4000 up to 4003 User Manual

Page 85: 4 initial, Operation, 1 attaching, The heating bath, 2 filling, Heating bath

Advertising
background image

Setup and initial use

1.02

Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control

83

4.4 Initial operation

4.4.1 Attaching the heating bath

Place the instrument base in a suitable location.

Fig. 4-1: Heating

bath

1. Place the feet of the heating bath in the tracks (1). Ensure that the controls

are facing forward.

2. If necessary, adjust the height of the feet.

– Loosen the feet by hand or with an open-ended wrench and adjust

them until the heating bath stands firmly without wobbling.

4.4.2 Filling the heating bath

The heating bath can be filled with a variety of liquid heating media.

When using to de-ionized or distilled water:

Add 0.2% borax (Na

2

B

4

O

7

*10 H

2

O) to the water.

The minimum/maximum volumes of the heating bath refer to the liquid level

with the evaporating flask immersed.

1. Select a liquid suitable for your application such as tap water, oil with a low

viscosity (40cP) and flash point (check safety data sheet) > 260 °C, water-

soluble polyethylene glycol.

2. Lower the evaporating flask.
3. Fill the heating bath with medium. The level must be between the

minimum and maximum marks.