Faults and troubleshooting

1.02

Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control

121

Fault/Message

Cause

Remedy

Main power switch turned off

Turn on main power switch

Main power switch defective

Contact service

Fuse defective

Replace fuse (see chapter

8.2, page 124)

Lift is in end position

Use other arrow button

Mechanical or motor defect

Contact service

Motorised lift does not work

Vertical stop not correctly set

Adjust height (see chapter

4.5.7, page 94)

VAC Senso T / Rotavac vario

control power not switched on

Turn power switch on

No evacuation

Fuse defective

Replace fuse (see chapter

8.2, page 124)

System not sealing properly

Check seals and connections

Check ground connections,

grease if required

Inadequate vacuum

Vacuum pump defective

Consult manufacturer's

documentation of vacuum

pump

Timer is programmed

Check or disable timer (see

chapter 6.5.2, page 116)

Instrument switches off

unexpectedly

(Laborota 4002 / 4003)

Fuse defective

Replace fuse (see chapter

8.2, page 124)

Instrument switches off

unexpectedly

(Laborota 4002 / 4003),

Display shows SAFE

Emergency shutdown of

Laborota 4002 / 4003 control

when a pressure of 1100 mbar is

detected

Press any button to restore

the Laborota 4002 / 4003

control to its initial state

Calibrate vacuum sensor

(see chapter 8.1.1, page 122)

Table 7-1: Troubleshooting