7 faults and troubleshooting, 7 faults, Troubleshooting – Heidolph LABOROTA 4000 up to 4003 User Manual

Page 122

Advertising
background image

Faults and troubleshooting

120

Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control

1.02

7 Faults and troubleshooting

Fault/Message

Cause

Remedy

Power cable not connected to an

AC outlet.

Plug power cable into an AC

outlet.

Instrument can not be switched

on.

Fuse defective

Replace fuse (see chapter

8.2, page 124)

Main power switch turned off

Turn on main power switch

Fuse defective

Replace fuse (see chapter

8.2, page 124)

Main power switch defective

Contact service

Heating bath connector cable is

not plugged in

(LABOROTA 4002 / 4003 control

only)

Plug in heating bath

connector cable

Heating bath heater defective

Contact service

Set Bath temperature higher than

heating bath control

Use heating bath control to

set temperature 20°C higher

than Set Bath temperature

No heating function

Temperature limiter tripped

If medium in heating bath: let

heating bath cool and reset

temperature limiter

If no medium in heating bath:

contact service

Main power switch defective

Contact service

Speed control is set to left stop.

Turn speed control to right

Drive defective

Contact service

Drive does not turn

Fuse defective

Replace fuse (see chapter

8.2, page 124)

Message in heating bath display:

Er1

Heating bath defective

Contact service

Message in heating bath display:

Er2

Heating bath defective

Contact service

Message in heating bath display:

Er3

Heating bath heated without

liquid heating medium

Switch instrument off

Check whether temperature

limiter has tripped. If so, reset

(see chapter 8.1.2, page 123)

Add liquid heating medium