Setup and initial use

1.02

Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control

85

Fig. 4-2: Connections

1 heating bath connector

cable

4 temperature sensor/boiling

temperature sensor

(Laborota 4010 / 4011 / 4002 /

4003)

2 power connection

5 heating bath control cable

(Laborota 4002 / 4003)

3 remote

control

(Laborota

4001 / 4011 / 4003)

6 VAC senso T / Rotavac vario

control (Laborota 4002 / 4003)

1. Insert the heating bath connector cable to the heating bath and the

appropriate socket of the instrument base.

Laborota 4002 / 4003 control:
2. Connect the bayonet plug of the heating bath control cable (5-pin) to the

heating bath and instrument base:
– Grasp the plug by the knurled edge and turn it counter-clockwise to

the stop.

– Press the plug into the socket in such a manner that the locking pin fits

into the groove.

– Turn the knurled ring clockwise until the plug clicks into place.

The bayonet plug is now locked onto the connector.
3. Plug the power cable into the power input socket on the rear of the

instrument.

4. Plug the power cable into an AC outlet.

4.4.5 Using the lift

The height of the drive can be adjusted in order to control the immersion depth

of the evaporating flask.

The flask may not contact the bottom or side of the heating bath while

immersed.