Heidolph LABOROTA 4000 up to 4003 User Manual

Page 93

Advertising
background image

Setup and initial use

1.02

Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control

91

2. Version

glassware G4

:

– Slide screw coupling (2) over the junction flange.
– Slide tension spring (3) over the junction flange.
– Fasten junction to the drive (4) by tightening the coupling (2) hand-

tight.

– Insert condenser in the upper receptacle of the junction.

3. Version

glassware G5 (cooling water independent)

:

– Slide screw coupling (2) over the cooling trap jacket flange.
– Slide tension spring (3) over the cooling trap jacket flange.
– Fasten cooling trap jacket to the drive by tightening the coupling hand-

tight.

– Insert the seal into the sealing groove of the cooling trap.
– Insert PTFE centring ring in the cooling trap and install the cooling trap

insert.

The vacuum causes the seal to be pressed into place and the system is

sealed hermetically by the inserted seal.
4. Version glassware G6 (condenser for reflux distillation):

Fig. 4-9:

Glassware G6

– Slide screw coupling (2) over the condenser flange.
– Slide tension spring (3) over the condenser flange.
– Fasten the screw coupling G6 (6) on the drive head (Fig. 4-8 (5)).
– Insert connector part G6 (5) with the seal G6 (4) in the screw

coupling G6 (6).

– Fasten junction (1) to the screw coupling G6 (6) using the screw

coupling (2).

– Insert condenser in the upper receptacle of the junction (1).

Advertising