Locate the battery cable fuse.

7

Locate the battery cable fuse positions.

8

Disconnect the battery cable fuse connectors in Position 1, Position 2, and Position 3.

9

Remove the used battery cable fuse. Locate Plug A, Plug B, and Plug C.

NOTE: The battery cable fuse contains two Plug A connectors, two Plug B connectors,

and one Plug C connector.

10

On the replacement battery cable fuse, connect both Plug A connectors to Position 1.

Table 23-1 Replacing the battery cable fuse on a portable instrument (continued)

Step

Description

Fuse, 3AG Slow-blow,

10 A 32 V Type T

Position 3

Position 2

Position 1

Plug A

Plug B

Plug C