INFICON 3000 Micro GC Gas Analyzer Reference Guide User Manual
Page 106
Advertising
23 - 2
PN
07
4-
55
8-
P1
A
3000 Micro GC Guides
6
Locate the battery cable fuse.
7
Locate the battery cable fuse positions.
8
Disconnect the battery cable fuse connectors in Position 1, Position 2, and Position 3.
9
Remove the used battery cable fuse. Locate Plug A, Plug B, and Plug C.
NOTE: The battery cable fuse contains two Plug A connectors, two Plug B connectors,
and one Plug C connector.
10
On the replacement battery cable fuse, connect both Plug A connectors to Position 1.
Table 23-1 Replacing the battery cable fuse on a portable instrument (continued)
Step
Description
Fuse, 3AG Slow-blow,
10 A 32 V Type T
Position 3
Position 2
Position 1
Plug A
Plug B
Plug C