Guide 9 integrating chromatograms, 3000 micro gc guide 9 – INFICON 3000 Micro GC Gas Analyzer Reference Guide User Manual
Page 39
9 - 1
PN
07
4-
55
8-
P1
A
3000 Micro GC Guides
3000 Micro GC Guide 9
Integrating Chromatograms
Table 9-1 Integrating Chromatograms
Step
Description
1
Integration events are used to customize the integration of certain peaks or regions of the
chromatogram. Refer to the 074-538 EZ IQ Operating Manual or online help in the EZ IQ
software for descriptions of individual events.
In EZ IQ, click Method >> Integration Events.
2
Select a channel to add or edit integration events. To change the channel, select the appropriate
channel from the drop-down list on the toolbar.
3
The Integration Events window will display the default integration events.
4
To add a new event, click on a blank field under Event in the table. Select the desired integration
event from the drop-down list.
5
Adjust the integration Start Time, Stop Time, and Value as desired.
6
Click Analysis >> Analyze to apply the changes.
7
To save the method, click File >> Method >> Save.