Integrating Chromatograms

Table 9-1 Integrating Chromatograms

Step

Description

1

Integration events are used to customize the integration of certain peaks or regions of the
chromatogram. Refer to the 074-538 EZ IQ Operating Manual or online help in the EZ IQ
software for descriptions of individual events.

In EZ IQ, click Method >> Integration Events.

2

Select a channel to add or edit integration events. To change the channel, select the appropriate
channel from the drop-down list on the toolbar.

3

The Integration Events window will display the default integration events.

4

To add a new event, click on a blank field under Event in the table. Select the desired integration
event from the drop-down list.

5

Adjust the integration Start Time, Stop Time, and Value as desired.

6

Click Analysis >> Analyze to apply the changes.

7

To save the method, click File >> Method >> Save.