INFICON 3000 Micro GC Gas Analyzer Reference Guide User Manual

Page 49

Advertising
background image

13 - 3

PN

07

4-

55

8-

P1

A

3000 Micro GC Guides

15

When finished, click Start.

16

When the analysis is complete, the chromatogram will be integrated, and the areas for
the peaks identified as calibration compounds will be entered into the method. In addition,
a calibration curve will be generated using these areas. The method is now calibrated for
a single level.

17

To save the calibration to the method, click File >> Method >> Save.

Table 13-1 Conduct a single level calibration (continued)

Step

Description

Advertising