When finished, click Start.
16
When the analysis is complete, the chromatogram will be integrated, and the areas for
the peaks identified as calibration compounds will be entered into the method. In addition,
a calibration curve will be generated using these areas. The method is now calibrated for
a single level.
17
To save the calibration to the method, click File >> Method >> Save.