Disconnect the communication cables leading from the GC modules, if present, to the
communications board connectors for the GC modules. If needed, also disconnect the power
cables leading to the fan.

8

Tilt the back of the GC module up until the mounting flange clears the alignment pins. Slide the
GC module towards the back of the instrument until it clears the inlet manifold frame and can
be lifted and removed from the front.

9

Inspect the gang block fitting on the bottom of the chassis to ensure the mating surface is clean.

Table 19-1 Changing a Performance Enhanced module (continued)

Step

Description

Back of GC

Remove cable

Remove

cables

Communication

board

Back view of GC modules

module

Tilt GC module

Mounting flange

Mounting flange

Tilt until mounting

flange is free from

aligning pins

2-channel 3000 Micro GC shown
Others are similar

Gang block, lower half